EXO3 Multiparameter Sonde

EXO3 Multiparameter Sonde 

5-Port Multiparameter Water Quality Sonde

Connectivity / Communications

Sonde: Bluetooth wireless technology, RS-485, and integral SDI-12 Output Options with Adapters:
USB, RS-232, or Modbus
Depth Rating 0 to 250 m (0 to 820 ft)
Desktop Software Compatible KorEXO
Flow Cell Yes
Medium

Fresh, sea or polluted water

Memory 512 MB total memory; >1,000,000 logged readings
Monitoring Yes
Multiparameter Yes
Operating Temperature -5 to 50 °C (23 to 122 °F)
Power 2 Alkaline Batteries
Sampling Yes
Size

Diameter: 7.62 cm (3.00 in)  Length: 58.67 cm (23.1 inc)

Smart Sensors / Ports 5 sensor ports (4 ports available when central wiper in use) Peripheral ports: 1 power communication port
Storage Temperature -20 to 80 °C (-4 to 176 °F). Exception: 0 to 60 °C (32 to 140 °F) for pH and pH/ORP sensors
User Calibratable Yes
Waterproof Yes
Weight 2.00 kg (4.41 lbs) with a full payload of 4 probes, 1 wiper, and probe guard installed

 

 


 

 

EXO Parameter MeasuredSensor**RangeAccuracyResponseResolution

 

Ammonium

(freshwater only)

 

Ammonium Sensor
Ammonia with pH sensor
SKU: 599710

0 to 200 mg/L (0 to 30˚C)

±10% of reading or 2 mg/L-N, w.i.g. T63<30 sec 0.01 mg/L

 

Barometric Pressure

Integral Barometer 375 to 825 mmHg ±1.5 mmHg from 0 to 50°C - 0.1 mmHg

 

Blue-green Algae,
Phycocyanin

 

Total Algae Sensor
SKU: 599102-01
0 to 100 μg/L;
0 to 100 RFU;
Linearity: R2 >0.999 for serial dilution of Rhodamine WT solution from 0 to 100 μg/mL BGA-PC equivalents T63<2 sec 0.01 μg/L;
0.01 RFU

 

Blue-green Algae,
Phycoerythrin

 

Total Algae Sensor
SKU: 599103-01
0 to 280 μg/L;
0 to 100 RFU;
Linearity: R2 >0.999 for serial dilution of Rhodamine WT solution from 0 to 280 μg/mL BGA-PE equivalents T63<2 sec 0.01 μg/L;
0.01 RFU

 

Chloride
(freshwater only)

 

Chloride Sensor
SKU: 599711
0 to 18000 mg/L-Cl (0 to 30˚C) ±15% of reading or 5 mg/L-Cl, w.i.g. T63<30 sec 0.01 mg/L

 

Chlorophyll

 

Total Algae Sensor

0 to 400 μg/L Chl;
0 to 100 RFU
Linearity: R2 >0.999 for serial dilution of Rhodamine WT solution from 0 to 400 μg/L Chl a equivalents T63<2 sec 0.01 μg/L Chl;
0.01 RFU

 

Conductivity1

 

 

Conductivity / Temperature Sensor
SKU: 599870

0 to 200 mS/cm 0 to 100: ±0.5% of reading or 0.001 mS/cm, w.i.g.;
100 to 200: ±1% of reading
T63<2 sec 0.0001 to 0.01 mS/cm (range dependent)

 

Depth - 10 m

 

Integral, Non-vented Depth Sensor3

0 to 10 m
(0 to 33 ft)
±0.04% FS (±0.004 m or ±0.013 ft) T63<2 sec 0.001 m (0.001 ft)
(auto-ranging)

 

Depth - 100 m

 

Integral, Non-vented Depth Sensor3 0 to 100 m
(0 to 328 ft)
±0.04% FS (±0.04 m or ±0.13 ft) T63<2 sec 0.001 m (0.001 ft)
(auto-ranging)

 

Depth - 250 m

 

Integral, Non-vented Depth Sensor3 0 to 250 m
(0 to 820 ft)
±0.04% FS (±0.10 m or ±0.33 ft) T63<2 sec 0.001 m (0.001 ft)
(auto-ranging)

 

Dissolved Oxygen, % air saturation

 

Optical Dissolved Oxygen Sensor
SKU: 599100-01
0 to 500%
air saturation
0 to 200%: ±1% of reading or 1% saturation, w.i.g.;
200 to 500%: ±5% of reading 4
T63<5 sec5 0.1% air saturation

 

Dissolved Oxygen, mg/L

 

Optical Dissolved Oxygen Sensor
SKU: 599100-01
0 to 50 mg/L 0 to 20 mg/L: ±0.1 mg/L or 1% of reading, w.i.g.;
20 to 50 mg/L: ±5% of reading 4
T63<5 sec5 0.01 mg/L

 

fDOM
(CDOM)

 

fDOM Sensor
SKU: 599104-01
0 to 300 ppb Quinine Sulfate equivalents (QSU) Linearity: R2 > 0.999 for serial dilution of 300 ppb QS solution
Detection Limit: 0.07 ppb QSU
T63<2 sec 0.01 ppb QSU

 

Level, Vented - 10 m

 

Integral Vented Level Sensor 0 to 10 m (0 to 33 ft) ±0.03% FS (±0.003 m or ±0.010 ft) T63<2 sec 0.001 m (0.001 ft)

 

Nitrate
(freshwater only)

 

Nitrate Sensor
SKU: 599709
0 to 200 mg/L-N (0 to 30˚C) ±10% of reading or 2 mg/L-N, w.i.g. T63<30 sec 0.01 mg/L

 

pH

 

 

 

 

pH Sensor
SKU:599701 guarded599702 unguarded

pH/ORP Sensor
SKU:599705 guarded599706 unguarded
0 to 14 units ±0.1 pH units within ±10˚C of calibration temp;
±0.2 pH units for entire temp range7
T63<3 sec8 0.01 units

 

ORP

 

 

pH/ORP Sensor
SKU:599705 guarded
599706 unguarded
-999 to 999 mV ±20 mV in Redox standard solution T63<5 sec6 0.1 mV

Temperature

 

Conductivity / Temperature Sensor
SKU: 599870

-5 to 35°C
35 to 50°C
±0.01°C2
±0.05°C2
T63<1 sec 0.001 °C

Turbidity9

 

Turbidity Sensor

SKU: 599101-01

0 to 4000 FNU 0 to 999 FNU: 0.3 FNU or ±2% of reading, w.i.g.;
1000 to 4000 FNU: ±5% of reading10
T63<2 sec 0 to 999 FNU = 0.01 FNU;
1000 to 4000 FNU = 0.1 FNU

 

Salinity

 

Calculated from Conductivity and Temperature11

0 to 70 ppt ±1.0% of reading or 0.1 ppt, w.i.g. T63<2 sec 0.01 ppt

 

Specific Conductance

 

Calculated from Conductivity and Temperature11 0 to 200 mS/cm ±0.5% of reading or .001 mS/cm, w.i.g. - 0.001, 0.01, 0.1 mS/cm
(auto-scaling)

 

Total Dissolved Solids (TDS)

 

Calculated from Conductivity and Temperature11 0 to 100,000 mg/L
Cal constant range
0.30 to 1.00
(0.64 default)
Not Specified - Variable

 

Total Suspended Solids (TSS)

 

Calculated from Turbidity and user reference samples 0 to 1500 mg/L Not specified T63<2 sec Variable

**Specifications indicate typical performance and are subject to change. All sensors have a depth rating to 250 m (820 ft), except shallow and medium depth sensors, ammonium, chloride, and nitrate. Accuracy specification is attained immediately following calibration under controlled and stable environmental conditions. Performance in the natural environment may vary from quoted specification.
w.i.g. = whichever is greater

EXO sensors are not compatible with YSI 6-Series sondes, sensors, or handheld.

 

 

 

 

Ordering Code!


Please contact us …

 

Visitors: 249,840