EXO3 Multiparameter Sonde
EXO3 Multiparameter Sonde
5-Port Multiparameter Water Quality Sonde
Connectivity / Communications |
Sonde: Bluetooth wireless technology, RS-485, and integral SDI-12 Output Options with Adapters: USB, RS-232, or Modbus |
---|---|
Depth Rating | 0 to 250 m (0 to 820 ft) |
Desktop Software Compatible | KorEXO |
Flow Cell | Yes |
Medium |
Fresh, sea or polluted water |
Memory | 512 MB total memory; >1,000,000 logged readings |
Monitoring | Yes |
Multiparameter | Yes |
Operating Temperature | -5 to 50 °C (23 to 122 °F) |
Power | 2 Alkaline Batteries |
Sampling | Yes |
Size |
Diameter: 7.62 cm (3.00 in) Length: 58.67 cm (23.1 inc) |
Smart Sensors / Ports | 5 sensor ports (4 ports available when central wiper in use) Peripheral ports: 1 power communication port |
Storage Temperature | -20 to 80 °C (-4 to 176 °F). Exception: 0 to 60 °C (32 to 140 °F) for pH and pH/ORP sensors |
User Calibratable | Yes |
Waterproof | Yes |
Weight | 2.00 kg (4.41 lbs) with a full payload of 4 probes, 1 wiper, and probe guard installed |
EXO Parameter Measured | Sensor** | Range | Accuracy | Response | Resolution |
---|---|---|---|---|---|
Ammonium (freshwater only)
|
Ammonium Sensor |
0 to 200 mg/L (0 to 30˚C) |
±10% of reading or 2 mg/L-N, w.i.g. | T63<30 sec | 0.01 mg/L |
Barometric Pressure |
Integral Barometer | 375 to 825 mmHg | ±1.5 mmHg from 0 to 50°C | - | 0.1 mmHg |
Blue-green Algae,
|
Total Algae Sensor
SKU: 599102-01
|
0 to 100 μg/L; 0 to 100 RFU; |
Linearity: R2 >0.999 for serial dilution of Rhodamine WT solution from 0 to 100 μg/mL BGA-PC equivalents | T63<2 sec | 0.01 μg/L; 0.01 RFU |
Blue-green Algae,
|
Total Algae Sensor
SKU: 599103-01
|
0 to 280 μg/L; 0 to 100 RFU; |
Linearity: R2 >0.999 for serial dilution of Rhodamine WT solution from 0 to 280 μg/mL BGA-PE equivalents | T63<2 sec | 0.01 μg/L; 0.01 RFU |
Chloride
|
Chloride Sensor SKU: 599711 |
0 to 18000 mg/L-Cl (0 to 30˚C) | ±15% of reading or 5 mg/L-Cl, w.i.g. | T63<30 sec | 0.01 mg/L |
Chlorophyll
|
Total Algae Sensor |
0 to 400 μg/L Chl; 0 to 100 RFU |
Linearity: R2 >0.999 for serial dilution of Rhodamine WT solution from 0 to 400 μg/L Chl a equivalents | T63<2 sec | 0.01 μg/L Chl; 0.01 RFU |
Conductivity1
|
Conductivity / Temperature Sensor |
0 to 200 mS/cm | 0 to 100: ±0.5% of reading or 0.001 mS/cm, w.i.g.; 100 to 200: ±1% of reading |
T63<2 sec | 0.0001 to 0.01 mS/cm (range dependent) |
Depth - 10 m
|
Integral, Non-vented Depth Sensor3 |
0 to 10 m (0 to 33 ft) |
±0.04% FS (±0.004 m or ±0.013 ft) | T63<2 sec | 0.001 m (0.001 ft) (auto-ranging) |
Depth - 100 m
|
Integral, Non-vented Depth Sensor3 | 0 to 100 m (0 to 328 ft) |
±0.04% FS (±0.04 m or ±0.13 ft) | T63<2 sec | 0.001 m (0.001 ft) (auto-ranging) |
Depth - 250 m
|
Integral, Non-vented Depth Sensor3 | 0 to 250 m (0 to 820 ft) |
±0.04% FS (±0.10 m or ±0.33 ft) | T63<2 sec | 0.001 m (0.001 ft) (auto-ranging) |
Dissolved Oxygen, % air saturation
|
Optical Dissolved Oxygen Sensor SKU: 599100-01 |
0 to 500% air saturation |
0 to 200%: ±1% of reading or 1% saturation, w.i.g.; 200 to 500%: ±5% of reading 4 |
T63<5 sec5 | 0.1% air saturation |
Dissolved Oxygen, mg/L
|
Optical Dissolved Oxygen Sensor SKU: 599100-01 |
0 to 50 mg/L | 0 to 20 mg/L: ±0.1 mg/L or 1% of reading, w.i.g.; 20 to 50 mg/L: ±5% of reading 4 |
T63<5 sec5 | 0.01 mg/L |
fDOM
|
fDOM Sensor SKU: 599104-01 |
0 to 300 ppb Quinine Sulfate equivalents (QSU) | Linearity: R2 > 0.999 for serial dilution of 300 ppb QS solution Detection Limit: 0.07 ppb QSU |
T63<2 sec | 0.01 ppb QSU |
Level, Vented - 10 m
|
Integral Vented Level Sensor | 0 to 10 m (0 to 33 ft) | ±0.03% FS (±0.003 m or ±0.010 ft) | T63<2 sec | 0.001 m (0.001 ft) |
Nitrate
|
Nitrate Sensor SKU: 599709 |
0 to 200 mg/L-N (0 to 30˚C) | ±10% of reading or 2 mg/L-N, w.i.g. | T63<30 sec | 0.01 mg/L |
pH
|
pH Sensor SKU:599701 guarded, 599702 unguarded pH/ORP Sensor SKU:599705 guarded, 599706 unguarded |
0 to 14 units | ±0.1 pH units within ±10˚C of calibration temp; ±0.2 pH units for entire temp range7 |
T63<3 sec8 | 0.01 units |
ORP
|
pH/ORP Sensor SKU:599705 guarded, 599706 unguarded |
-999 to 999 mV | ±20 mV in Redox standard solution | T63<5 sec6 | 0.1 mV |
Temperature |
Conductivity / Temperature Sensor |
-5 to 35°C 35 to 50°C |
±0.01°C2 ±0.05°C2 |
T63<1 sec | 0.001 °C |
Turbidity9
|
Turbidity Sensor SKU: 599101-01 |
0 to 4000 FNU | 0 to 999 FNU: 0.3 FNU or ±2% of reading, w.i.g.; 1000 to 4000 FNU: ±5% of reading10 |
T63<2 sec | 0 to 999 FNU = 0.01 FNU; 1000 to 4000 FNU = 0.1 FNU |
Salinity
|
Calculated from Conductivity and Temperature11 |
0 to 70 ppt | ±1.0% of reading or 0.1 ppt, w.i.g. | T63<2 sec | 0.01 ppt |
Specific Conductance
|
Calculated from Conductivity and Temperature11 | 0 to 200 mS/cm | ±0.5% of reading or .001 mS/cm, w.i.g. | - | 0.001, 0.01, 0.1 mS/cm (auto-scaling) |
Total Dissolved Solids (TDS)
|
Calculated from Conductivity and Temperature11 | 0 to 100,000 mg/L Cal constant range 0.30 to 1.00 (0.64 default) |
Not Specified | - | Variable |
Total Suspended Solids (TSS)
|
Calculated from Turbidity and user reference samples | 0 to 1500 mg/L | Not specified | T63<2 sec | Variable |
**Specifications indicate typical performance and are subject to change. All sensors have a depth rating to 250 m (820 ft), except shallow and medium depth sensors, ammonium, chloride, and nitrate. Accuracy specification is attained immediately following calibration under controlled and stable environmental conditions. Performance in the natural environment may vary from quoted specification.
w.i.g. = whichever is greater
EXO sensors are not compatible with YSI 6-Series sondes, sensors, or handheld.
Please contact us …